Online UV Dissolved Ozone Analyzer
DK-27TR
Overview
The DK-27TR Trace Ozone Analyzer is a special ozone measuringinstrument that uses patented light absorption technology to achieve high precision, high stability, ppb concentration measurement, high reliability and minimum maintenance for the initial Design concept,analytical instrument electrodeless consumables and membrane consumables, no moving parts loss, the only maintenance requirement is to replace the UV lamp every few years, the key industrial advantages include ultra-sensitivity, no drift stability and no other gas cross interference.
The DK-27TR trace ozone analyzer controller is based on a micro-processing core including analog and digital interfaces, and three available alarm relays.
Characteristics
l ppb level dissolved ozone
l Patented ultraviolet absorption measurement technology
l No electrode and no film consumables
l No moving parts without long-term maintenance
l High stability measurement
l Immune cross interference
l Analog and digital interfaces
l Convenient unattended operation
Application
l Measurement of trace dissolved ozone
l Ultra-purity Water System
l American Pharmacopoeia pure water system
l Ozone injection after end
l UV irradiation after end
l
Other applications
Industr
Industries
l Semiconductor
l Pharmacy
l Bioengineering
l Laboratory research
l Other
Specification
Sensor | |
Application | Trace dissolved ozone analyzer |
Measuring principle | Ultraviolet Absorption, Dual Optical Path Sampling (Patent) |
Ultraviolet light source | Low pressure mercury lamp |
Unit of measurement | ppbW, ppmW, mg/ |
Range | 0-1000 ppbW (more customizable) |
Repeatability | 3 ppbW or 2% readings (higher performance customizable) |
Flow rate | 0.75~1.5 SLPM |
Pressure | 15 ~ 20 PSI |
Sampling connection | 1/4” Swagelok™ (other brands are also available) |
Shell | NEMA 4X/IP65 |
Size | 6.3”W x 15.9 "H x 3.9" D (160 x 405 x 99 mm) |
Maintenance | Replace the UV lamp every 24~36 months (no other maintenance) |
Controller | |
Display | 16 characters, numbers, white LCD on a black background |
Resolution | 1ppbW |
Period | Continuous measurement every 0.5 seconds |
Analog output | Photoelectric Isolation 4-20 mA or 0-10V dc |
Digital interface | RS-232C half-duplex |
Diagnosis | Built-in diagnosis, instrument error relay |
Alarm relay | 3 Programmable DPDT alarm relays |
Power Supply | 90-240 VAC, 50/60 Hz |
Install | Wall mounted NEMA4 certified enclosure |
Size | 10”W x 8.5 "H x 5.6" D (254 x 216 x 143 mm) |
Certification | CE |