LP1 laser photoacoustic spectroscopy gas analyzer
Overview
LP1 laser photoacoustic spectrum gas analyzer integrates a semiconductor laser source with high precision and high selectivity tunable controllable spectrum, and a cantilever beam enhanced photoacoustic spectrum technology with wide dynamic range. The zero background noise technology provides high stability without repeated calibration for several months, which makes LP1 an ideal choice for the trace concentration measurement requirements of PPB and PPT.
High sensitivity and a variety of laser light source selection
The photoacoustic spectrum instrument works in an unresonant mode. The signals from two laser sources can be measured simultaneously by using different modulation modes. The resulting photoacoustic signal equivalent quantifies directly the concentrations of different measured gases in a photoacoustic sample chamber.
The modulation wavelength of the laser and the use of low voltage measurement make the analysis of different mixed gases have excellent selectivity. Based on this background noise signal minimization technique, the results of the measurement can be shifted slightly and the calibration period can be extended.
For each suitable laser source, LP1 can be equipped with a wide range of NIR near infrared semiconductor lasers ,and continuously increase the optional laser. Such as quantum cascade lasers (QCL), or near infrared optical parametric oscillators (IR OPO), Continuously increase the gas measurement range and enhance the excellent function of the instrument.
Features
Configurable a variety of laser light according to the operation conditions.
Simultaneous analysis of multicomponent gases
Ppb level measurement
High selectivity
Low sample volume & wide dynamic range
Long calibration period
Free maintenance
Built-in 2 point sampling
Friendly calibration interface
Temperature and pressure fluctuation compensation
Multifunction analysis task
Operation interface pictorial
Results illustrated with pictures and texts
No consumables & no moving parts
Application
Gas impurities
Greenhouse gas (GHG)
Air quality
Occupational safety
Process control
Automated production
Refrigerant leakage
Exhale analysis
Security screening
Natural gas and synthetic gas
Soil analysis
Technology
Semiconductor laser based on photoacoustic spectrum with enhanced cantilever optical microphone
Sample chamber with constant temperature of 50℃
Patented ultra-sensitive optical microphone, based on MEMS cantilever sensor
19”3U industrial housing
Industrial computer with 5.7 "color display
Settable alarm limited detection for the analyzed gas
2GB built-in storage, sufficient for more than 1 year for continuous measurement at the shortest sampling interval.
Connect USB port, Ethernet network port and serial port through USB or PC for data transmission
Four gas inlets on the back of the instrument, two customizable as filter device and purification device
Automatically compensates for temperature and pressure fluctuations
Specification
Dimension : 48.4cm W x 13.9cm H x 44cm D (19,1in W x 5,5in H x 17.3in D)
Weight : approximately 13kg
Total sampling volume: 30ml
Pneumatic connection: 6/4mm tube
Electrical connections:
Input voltage: 100~240Vac, 50~60Hz
Input power: 100W
Communication interface: Ethernet,USB,RS232,RS485
Condition
Operation condition:
Temperature: 0°C ~ +45°C
Humidity:<90% RH, no condensation
Pressure: atmospheric pressure
Protection grade: IP20 (IEC 529)
Storage condition:
Temperature range: -20°C ~ +60°C
Sampling condition:
Temperature : 0 ~ +49°C, no condensation
Pressure: 930mbar ~ 1100mbar
Flow rate: approximately 1L/min
Particle: < 1 μm
Features
Response time: typical 5s, customizable
Limited detection : typical ppb level customizable
Dynamic range: Typical 5 orders of magnitude
Repeatability:<1% measurement value
Accuracy : 2% measurement value
Temperature stability: Automatic temperature compensation
Pressure stability: Automatic pressure compensation