EDK 7100 online ozone gas analyzer
Overview
Online EDK 7100 ozone gas analyzer of ETG Italy, based on Beer-Lambert Law for ultraviolet light sources. Ultraviolet light at the wavelength of 254nm is absorbed by O3 when passing through the sample chamber of O3 gas. The absorption value and the concentration of O3 follows Beer-Lambert Law. By measuring the change value of ultraviolet light signal before and after absorption and comparing with the reference light source, the concentration value of O3 was analyzed.
Characteristics
Second level response, quick measurement
Optional multipoint sampling (MPS)
Modular design, simple maintenance
High stability optical path system and UV light source
Plug and play online rack type 3U chassis
Double light path nondispersive ultraviolet absorption spectroscopy
Good repeatability (2%), high reliability (IP54)
Automatic temperature and pressure compensation (built-in pressure and temperature sensor)
Simple operation (5.7 "touch screen), strong, reliable and durable (carbon steel case)
Application
Environmental gas, clean room, pollution source, meteorological climate, ozone preparation, petrochemical raw material gas, rubber materials, papermaking, textile, pharmaceutical, essence and flavor, water treatment, medical and health, food and other industries in the process of scientific research and production of ozone monitoring and analysis
Specification
Range : 0~200ppb、0~10ppm、0~200ppm、0~1000ppm、0~10%,customizable
Precision/repeatability : <±2% F.S.
Accuracy: <±2% F.S.
Zero drift: Monthly less than 0.2% (absolute value) (nitrogen is zero)
Range calibration: Factory calibration, third-party authority calibration
Response Time: T90<2s
Warm up time: <10 minutes
Technology: Nondispersive ultraviolet absorption spectroscopy NDUV
Flow rate: 0.4LPM
Pressure: 0-5 psig
Sample temperature: 0-50℃
Sample connector: 1/4” Polyvinylidene fluoride Swagelok fittings.