• Photoacoustic Gas Analyzer
  • Dissolved O2, O3, H2 analyzer
  • Sonic densitometer
  • Dissolved carbon dioxide analy
  • Other instrumentation
  • M100 SYN  SynGAS Portable MULTICOMPONENT ANALYZER

    M100 SYN  SynGAS Portable MULTICOMPONENT ANALYZER


     

     

    The MCA 100 Bio series of gas analyzers by ETG it’s the ideal solution for biogas measurement and industrial combustion applications because of their accuracy, stability, reliability, broad measurement range, and the variety of available form factors.


    Unlike other analyzers, ETG MCA 100 non-dispersive infrared (NDIR) gas analyzers measure multiple gases in an instrument with a single optical path platform.

     

    Single-gas analyzers are inadequate when using methane as a biofuel because the gas often contains large amounts of CO as a contaminant. ETG analyzers have the ability to measure CO2, CO, and O2 in addition to methane and therefore provide the optimal combination of gases for combustion optimization.

     

    The enhanced optics and electronics of our NDIR analyzers have virtually eliminated zero drift after the initial warm up period. The temperature and pressure compensation eliminates the major causes of span drift in many NDIR instruments.

     

    M100 SYN BIO  MULTICOMPONENT ANALYZER FOR BIOGAS

    Measurement Method

    Gas

    Resolution

    Range

    Accuracy

    Precision

    Time

     

    NDIR 

    ( Non-Dispersive Infrared )

    Methane

    0.01%

     

    0-100%

    201ppm-40%

    40-80%

    80-100%

    +/- 1% F.S. 

    +/- 2% rel

    +/- 1% rel

    +/- 5% rel

    +/- 0.8%

    T90 & T10 <

    10 s

     

    NDIR 

    ( Non-Dispersive Infrared )

    Carbon

    Dioxide


    Carbon 

    Monoxid


    0.01%



    0.01%

    0-50%

    0.00-16.00%


    0.01-20.00%

    (up to 40%)


    +/- 0.8 F.S.

    +/- 0.3% abs 

    or +/- 3.0% rel

    +/- 5% rel

    +/-0.7%

    +/- 0.1% abs or +/- 0.8% rel

    T90 & T10 <

    10 s

    Electro-chemical sensor

    Oxygen

    0.01%

    1.01-25.00%

    +/- 0.1% abs  or +/- 3.0% rel

    +/- 0.1% abs or +/- 1.5% rel

    <40s from ambient to 0.15 O2

    Electro-chemical sensor

    Hydrogen 

    0.01%

    0-10%

    0-20%

    0-50%

    0-100%

    +/- 0.5  abs or +/- 3.0% rel

    +/- 0.5  abs or +/- 3.0% rel


    T90 & T10 <

    20 s

    Specifications

    Response Time 

    Response times are specified at  a sample flow rate of a 1 liter  per minute through   the MCA 100 sample cell

    Data Refresh Rate

    1 second

    Warm-up Time

    30 seconds ready,3 minutes useable,30 minutes to full performance

    Operating Temperature

    0°C to 70° C ( 32° to 158° F)

    Operating Humidity

    To 95% RH ( Non-condensing)

    Operating Altitude

    -300 to 3.000 m ( -1.000 to 10.000 ft )

    Communications 

    USB port ( standard ) 

    Case Protection

    IP67(1 meter submersion for 30 minutes)  • Def Stan 81-41/STANAG 4280 


    M100 SYN BIO  
    MULTICOMPONENT ANALYZER FOR BIOGAS

    Battery

    12 – 24  Vcc with Internal Charger and light charge indicator

    Monitor

    Touch Screen  5.7” Resistive type

    Calibration

    Zero & span user  selectable. Automatic calibration ( optional)

     External Electrical supply

    from 100 to 240 Vac 47-63 Hz

    Pneumatic connection

    Rapid fittings 6.0  OD 4.0  ID

    Mechanical dimensions

    50 x 28 x 40 cm 

    Weight

    7.5 kg

                    

                                    Charts and Figures                                                                    Trend Charts and Numbers


    All the ETG MCA 100 and 6500 series for Syngas have Remote Support and Remote Access

    · File transfer

    · Data monitoring by your PC , tablet  and smartphone

    · Software update


    Consulting phone 010-8243-3533
    Duke. All rights reserved Beijing tech technology Co. Ltd. 京ICP备14007870号-1